CVE-2014-5431

CVE-2014-5431

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Source: CVE-2014-5431

CVE-2014-5432

CVE-2014-5432

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Source: CVE-2014-5432

CVE-2014-5433

CVE-2014-5433

An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Source: CVE-2014-5433

CVE-2014-5434

CVE-2014-5434

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Source: CVE-2014-5434

CVE-2019-10063

CVE-2019-10063

Flatpak before 1.0.8, 1.1.x and 1.2.x before 1.2.4, and 1.3.x before 1.3.1 allows a sandbox bypass. Flatpak versions since 0.8.1 address CVE-2017-5226 by using a seccomp filter to prevent sandboxed apps from using the TIOCSTI ioctl, which could otherwise be used to inject commands into the controlling terminal so that they would be executed outside the sandbox after the sandboxed app exits. This fix was incomplete: on 64-bit platforms, the seccomp filter could be bypassed by an ioctl request number that has TIOCSTI in its 32 least significant bits and an arbitrary nonzero value in its 32 most significant bits, which the Linux kernel would treat as equivalent to TIOCSTI.

Source: CVE-2019-10063

CVE-2019-7715

CVE-2019-7715

An issue was discovered in the Interpeak IPCOMShell TELNET server on Green Hills INTEGRITY RTOS 5.0.4. The main shell handler function uses the value of the environment variable ipcom.shell.greeting as the first argument to printf(). Setting this variable using the sysvar command results in a user-controlled format string during login, resulting in an information leak of memory addresses.